The FDA-posted notice concerns specified LG3800-30VT devices used with Targeted VT enabled; patients are told to consult their care providers.

What the record says

React Health published a November 17 update to a correction for specified Luna G3 BPAP System devices, model LG3800-30VT, when Targeted VT is enabled. FDA posted the company notice November 19. It lists 7,533 distributed devices in the United States and Puerto Rico, with affected serial numbers in an appendix.

The company says a storage and transmission error can make historical pressure statistics inaccurate on the SD card or in React Health Connect. It describes a risk that clinicians could make decisions using incorrect historical data; the issue is not described as affecting users who have not enabled Targeted VT.

React Health reports 15 complaints and two serious adverse reports associated with the issue. It says firmware G3-2.00.80 is available through the company and equipment providers.

The patient instructions say to contact the healthcare provider or equipment supplier, continue prescribed therapy pending further guidance, and not independently change settings or discontinue respiratory therapy.

Source details and supporting locations.

Practical next step

Contact the prescribing clinician or equipment provider to check the serial number and coordinate correction. Follow the notice’s patient instructions rather than attempting clinical setting changes yourself.

Limits and what is not established

This summarizes a device-specific reporting correction, not an individual assessment of therapy. The complaint counts are company-reported; affected serial numbers and clinical decisions require provider review.

Sources checked November 22, 2026. Later notices may change instructions or arrangements.